Model | BIOBASE1001 | |
Sample& Reagent Unit | Sample Racks | 6*20 push-pull sample rack |
Sampling Channel | 1 channel | |
Liquid Level Detection | Air pressure level detection, clot detection function | |
TIP | 1000μl, 2 TIP racks | |
Pipetting | 10ul~1000ul | |
Pipetting Precision | 100μl: CV≤1%, Accuracy≤2.5% 1000μl: CV≤0.5%, Accuracy≤1% | |
Reagent Position | 2 reagent racks | |
Dilution Plate Position | 1 dilution plate position with incubation and shaking | |
Washing Unit | Washer | 1 unit |
Washing Head | 16-channel (32-pin), double row cleaning, hanging installation, with leakage recovery | |
Injection Uniformity | CV<1.5% | |
Washing Residual | <2μl | |
Reading Unit | Reading Channels | 8 independent photometric channels, mono and dichromatic reading |
Absorbance Range | 0.000~4.000Abs | |
Linear Range | 0.000~3.000Abs | |
Repeatability | CV≤1.0% | |
Stability | ≤±0.003Abs | |
Spectral Range | 400~700nm | |
Optical Filters | 4 standard filters (405, 450, 492, 630nm); Total 10 filters positions |