Model
|
BIOBASE2000
|
|
Sample pipetting unit
|
Sample rack
|
24 sample tubes/rack, at most 23 racks
|
Primary sample tubes
|
11.5-13mm
|
|
Pipetting
|
8μl~1000μl, (10μl) CV<5%
|
|
Time to dispense
|
12 minutes adding 96 samples
|
|
Dispensing system
|
1 aspirating-dispensing probe with X-Y-Z movement;dilution system with a syringe of 1000μl, continuous flow washing station.
|
|
Reagent pipetting unit
|
Reagent racks
|
2 reagent tanks(60ml)/ rack, at most 23 racks
|
QC racks
|
24 positions/QC rack, at most 23 racks
|
|
Pipetting
|
8μl~1000μl,1μl stepping
|
|
pipetting precision
|
(100μl )CV≤0.5%
|
|
Time to dispense
|
3 minutes(typical)^100μl of 96 reagents to plate from reagent bottle
|
|
Washing unit
|
Programmable
|
Yes(options & parameters)
|
Wash containers
|
3 wash buffers at 2.0L,1 wash buffers at 5.0L,with level-sensing
|
|
Waste containers
|
25L with waste fμll sensor
|
|
Washing system
|
two 8 nozzle manifolds, one to dispense and the other one to aspirate
|
|
Washing residual
|
<2μl
|
|
Reading unit
|
Reading system
|
8 independent photometric channels with mono and bichromatic reading capability
|
Photometric range
|
0.000 ~3.000 OD
|
|
Spectral range
|
400-700nm
|
|
Optical filters
|
2 precision filters (450and 630nm); 2 more on request (4 filters max, 405 nm, 450 nm, 492 nm and 630 nm)
|
|
Reading inaccuracy
|
inaccuracy±1% from 0.000 to 1.500 OD, ±2% from 1.500 to 3.000 OD
|
|
Incubation unit
|
Incubators
|
4 independent units
|
Programmable
|
Yes
|
|
Temperature range
|
RT to 45 °C
|
|
Temperature accuracy
|
±0.5ºC
|
|
Temperature staility
|
±0.5ºC
|
|
Software unit
|
Software system
|
Windows 7/8/107 or above, LIS system available
|
Function application
|
work list set up, patient reports and resμlt filing; calibration curve storage for each carried out test
|
|
Data reduction
|
cut-off (qualitative)
|
|
Others
|
Power Supply
|
AC 110/220V±10%,50/60Hz,450W
|
External Size(W*D*H)
|
1160*700*680mm
|
|
PackageSize(W*D*H)
|
1200*800*1110mm
|
|
Gross Weight
|
150kg
|